GC-4100 Series Gas Chromatograph
Features
The GC-4100 Series is NANBEI's new generation of gas chromatographs. As compared with the GC-4000A Series, a number of technical improvements has resulted in better performance and greater automation. Up to ten gas paths can be controlled using EPC flow or pressure control, including the carrier gas, ignition gas, auxiliary gas, split flow and makeup gas, among others. Detector and column oven temperatures are controlled by a new and improved PID method, with a control accuracy of ± 0.03 ℃.
● The new TCD has less thermally conductive volume, which improves stability time. Our TCD is constructed with anti-oxidant components and features a constant average temperature bridge power supply to heat the filament. This results in an increase in sensitivity while improving lifetime. Using split flow injection with a capillary column and TCD, six consecutive analysis of benzene in toluene results in an RSD of less than 2.5%, as well as peaks with a small FWHM.
● The volume of the micro ECD cell has been reduced, resulting in greater baseline stability. Capillary columns can now be used with ECD. The sensitivity has been improved significantly. The detection limit now reaches 2 × 10-14 g/mL (γ-666).
GC-4100 Gas Chromatograph Technical Parameter
1、Column oven
Dimension(mm) |
280mm×300mm×280mm(L×D×H) |
Oven Capicity |
280mm×190mm×280mm(L×D×H) |
Temperature Range |
(RT+5)℃~400℃,increments 0.1℃ |
Accuracy(Stability) |
≤0.1%(Or less than±0.03℃) |
Programmed temperature Rate |
0.1℃/min~40℃/min (increments 0.1℃/min) |
Programmed temperature Repeatability |
≤1% |
Programped temperature Order |
14 ramp |
Over temerature protection |
Dual over-temperature protection, set the protection temperature through the instrument control panel and set the overtemperature protection via the instrument circuit hardware |
2、Detector
Equipped with 3 detector at the same time at most.
Can equipped with TCD/FID/FPD/ECD/NPD etc.
TYPE | TCD | FID | FPD | ECD | NPD |
Sensitivity Detection |
S≥10000mV·mL/mg |
D≤1×10-11 g/s |
D≤8×10-11g/s(S);D≤2×10-12g/s (P) |
D≤3×10-14g/mL |
D≤1×10-12g/s (N); D≤1×10-12g/s (P) |
Baseline Drift |
≤0.2mV/30min |
≤3×10-14A/30min
(or ≤0.3mV/30min, 1010Ω) |
≤2×10-14A
(or≤0.2mV/30min) |
≤0.3mV/30min |
≤2×10-14A/30min
(or ≤0.2mV/30min,1010Ω) |
Baseline Noisy |
≤0.1mV |
≤1×10-14A (或≤0.1mV, 1010Ω) |
≤1×10-14A
(or≤0.1mV) |
≤0.1mV |
≤1×10-14A
(or ≤0.1mV,1010Ω) |
3、 Workstation/Software
The temperature of each module, can be set on the software, including the oven, the inlet, the detector and the EPC.
Signal Acquisition Frequency |
1~100Hz |
Output signal Range |
-1250 mV~1250 mV |
Acquisition Precision |
1μV |
Channels |
Up to 4 at the same time |
4 、Host parameter
Dimension: |
565mm×520mm×530mm |
Weight: |
60 kg |
Watts: |
2500 W |