Brand:NANBEI
Model:LHH-250GSP
Application:
LHH series medicine stability test box/general medicine stability test box
Summary:
Adopt balance temperature and humidity adjusting method, select imported unclosed industrial compressor, humidity sensor, touch-screen programmable controller (1-99 sections), recorder etc for deploy medicine stability test box, they are stable, safe and reliable
Features:
1 adopted 304 standard of stainless-steel chamber, semicircular arcs at comers for easy cleaning
2 the major parts such as temperature and humidity controller,compressor, circulating fan adopts import product
3 unique air duct circulating design, air ducts ensures wind distribute evenly in chamber, a test hole(diameter-25mm) is placed in left side of chamber
Specifications:
model | LHH series medicine test box | LHH series general medicine stability test box | ||||
LHH-150SD | LHH-250SD | LHH-150SDP | LHH-250SDP | LHH-150GSP | LHH-250GSP | |
Electrical requirements |
AC 220V 50HZ | |||||
Temperature range | 0C-65℃ | NO lighting:0C-65℃ lighting:10C-65℃ | ||||
Temperature control | ±0.5℃ | |||||
Humidity range | 40-95%RH | |||||
Humidity deviation | ±3%RH | |||||
Illumination range | N/A | 0-6000LX | ||||
controller | Special temperature and humidity digital display instrument | Imported touch-screen programmable controller | ||||
Interior dimension(mm) | 500*505*600 | 550*520*900 | 500*505*600 | 550*520*900 | 500*505*705 | 550*520*1020 |
Overall dimension(mm) | 630*705*1150 | 680*720*1540 | 630*705*1150 | 680*720*1540 | 630*705*1270 | 680*720*1660 |
Timing range | 3(pcs) |
Satisfying the ICH2003 Q1A(2)guiding principle /GMP2005 edition Chinese pharmacopoeia stability test conditions:
1 short-time test:40C±2C/75%RH±5%RH
2intermediate qualification:30c±2C/65%RH±5%RH
3 long-run test:25C±2C/60%RH±5%RH or 30C±2C/65%RH±5%RH
Execution standard:2005 edition pharmacopoeia medicine stability test guideline and relative items in GB 10586-2006