5L High temperature circulator
5L High temperature circulator

5L High temperature circulator

Brand:NANBEI

Model:5L

Application:

product details

 High temperature circulation 

Introduction:GX series high temperature circulator is our design of a high temperature heat source, suitable for jacketed reactor, chemical reaction, high temperature distillation, semiconductor industry. GX series high temperature circulating device made up of domestic defects of similar products, price is greatly lower than imported products.

Feature:

1 High temperature circulator is suitable for fermenter,jacketed glass reactor and other things.

2 Outside circulation could rise the temperature in the jacket,therefore materials in which would react.

3 Temperature fluctuation is in ±0.2°C.

4 Resolution of digital display is 0.1°C.

5 Flow of outer circulation is 7~15L/min.

6 Auto-alarm of over temperature.

7 Stop heating when the temperature is over.

8 There are six models that can be chose:5L,10L,15L,20L,30L,50L.

9 Our high temperature adopts the newest control technique,easy to operate,great

stability and high precise.

10 The PID can be set by yourself.

11 LED double window respectively digital temperature measurement and temperature setting value.

12 Microcomputer intelligent could modify the error of measuring value.

Operation steps:

1 Pour liquid into the sink,the distance between liquid level and the working platen should not less than 20mm.

2 Chose oil for the liquid.The initial temperature of the oil must higher than working temperature, range of which is more than 15°C.

3 Connect the out circulator pump with external equipment,which includes jacketed reactor, chemical reaction, high temperature distillation or other equipment.First connect outlet pipe with the inlet pipe of external equipment,then connect inlet pipe of the circulator pump with the outlet pipe of the external equipment.

4 Turn on the power,then switch the“POWER”button and the”CIRCULATOR”button on.

Note:

There must be liquid in the tank before switch the “POWER”button.

The power supply should be AC 220V/50Hz,and it should be connected with ground.

3.This high temperature circulator should be placed at dry and ventilated places,preventing heat and damps.There should be no objects around its back and sides in 300mm.

4 If you do not use it temporary,please turn off all buttons and the power.

Parameters:

Model

Volume(L)

Temperature(C)

Temperature   fluctuations(C)

Flow(L/min)

Tank   size(mm)

Tank   Depth(mm)

GX-2005

5

Normal-300

±0.2

8

240*150*150

150

GX-2010

10

10

280*190*200

200

GX-2015

15

15

280*250*200

200

GX-2020

20

15

280*250*280

280

GX-2030

30

15

400*330*230

230

GX-2050

50

15

500*330*300

300

             