NDK200-1A Sample Concentrator:
Application:
NDK200-1B visual nitrogen concentrator is mainly used for enrichment of large-volume sample or preparation
(such as drug screening, hormone analysis, liquid phase, and mass spectrometry in the analysis of sample preparation).
Working principle: the nitrogen into the heated surface of the sample and make the solvent t evaporate and separated rapidly,
thereby reaching the sample enrichment purpose without oxygen, keep the sample purer. Nitrogen concentrator instrument can
concentrated dozens of samples simultaneously, saving time and high efficient, instead of rotary evaporation instrument.
This type of product has a unique feature: The liquid level position of sample concentration can be observed at any time.
Product features:
1, the heater allows samples to be heated rapidly to the evaporation temperature, through the gas blow to the solution surface
via needle at the same time, prompting rapid evaporation and concentration solution.
2, Gas needle position in the gas chamber can be changed, suitable for various kinds of tubes.
3, the gas chamber height can be adjusted, the length of standard gas needle is 80 mm.
4, each blow needle can be controlled independently, can be blown, separate flow regulation, no waste of gas.
5, the entire system can be arranged in the ventilation cabinet when process concentration of toxic solvents,
6, Built in over-temperature protection device, automatic fault detection and alarm function.
7, LED display, temperature and time diminishing display, the operation is simple and convenient.
8 configured standard gas chamber and special adjustable bracket.
Technical Parameters:
Model | NDK200-1A 96 holes nitrogen concentrator |
Temperature range | RT+5℃ to 150℃ |
Heating time | ≤30min (从40℃升至150℃) |
Temperature accuracy(40~100℃) | ±0.5℃ |
Temperature accuracy(100~150℃) | ±1℃ |
Temperature uniformity(100℃) | ±0.5℃ |
Blocks Quantity | 1 pieces |
Time range | 0.1 |
Nitrogen flow | 99h59min |
Nitrogen pressure | 0—10L/min |
power | ≤0.1MPa |
Dimensions | 200W |
Blocks Selection | 280x220x400 mm |
Temperature range | Refer to the “blocks replacement for NDK-1A Series” |
Conventional configuration
The user can select the blocks from DT16 and DT17, blocks can be interchanged, and conventional configuration is DT06 blocks.
Blocks replacement for NDK-1A Series
Type | Tube diameter | Tubes | Block size (mm) |
DT16 | 0.2ml centrifugal tube | 96 | 78 x 114 x 26 |
DT17 | Plate(no holes) | 96 hole enzyme plates | 81 x 123 x 19 |
DT16, DT17 is special configured for NDK200-1A