NP Series Laboratory Ultrapure Water System
NP Series Laboratory Ultrapure Water System

NP Series Laboratory Ultrapure Water System

Brand:NANBEI

Model:

Application:

product details

 Laboratory Ultrapure Water System Series NP


2010-7-6

 

ultra pure water system

Desk Type 

ultrapure water machine

 Standing Type

 

Feature & Function:
1 Full automatic and continuous water production, automatic display of working status, automatic protection of system.
2  Self inspection while starting up, automatic stop when reservoir is full of pure water, protection alarm when lack of feed water, automatic reposition when power is off.
3  System is no heating, no noisy when working, low power.
4  Online monitor display of ultrapure water quality and temperature on LCD screen.
5  Imported UF parts can guarantee effective elimination of pyrogen and water flow.
6  UV sterilizer with long lifetime and built-in six circulation design can guarantee high effective elimination of TOC.
7  RO membrane flushing can prolong lifetime of consumables and accessories.
 
Guide for Order:
1.   For special configuration and optional function, add “T”at the front of standard model code.
2.  For feed water is distilled water or pure water, add “V”at the back of standard model code.
3.  For deep treatment function, add “S”at the back of standard model code.
 
 
Model
NP 05/ 10/ 20/ 40/ 60-Y 
NP60-CY
Feed Water
City tap water,  feed pressure: 0.15-0.4Mpa,
temperature: 5-40℃
RO water, de-ionized water,
distilled water
Flow rate of
water production
5 / 10 / 20 / 40 / 60 L/hr
60 L/hr
Flow rate of dispense water
1.5-2L/min
Output water parameter
Pure water: quality confirms to the GB6682-2000 criteria of type Ⅲ water in laboratory application.
Ultrapure WaterResistivity: 18.2 MΩ.cm@25℃, Pyrogen<0.001EU/ml, TOC:1-3ppb, Microbe<1cfu/ml, Particle (>0.01 um)<1/ml
Application Range
Molecular Biology, Life Science, Genetic Research, Tissue Culture, Tube Baby, Bio-engineering, Animal Cell Culture, Culture Medium Preparation, DNA sequencing, Amino Acid Analysis, Protein Purification, etc.
Power
220V/50Hz;  50-100W
External Dimension
500×310×400 / 450×300×400 mm
450×300×400 mm
 
Model
P  05/ 10/ 20/ 40/ 60-W
P60-CW
Feed Water
City tap water,  feed pressure: 0.15-0.4Mpa,
temperature: 5-40℃
RO water, de-ionized water,
distilled water
Flow rate of
water production
5 / 10 / 20 / 40 / 60 L/hr
60 L/hr
Flow rate of dispense water
1.5-2L/min
Output water parameter
Pure water: Resistivity ≥0.2MΩ.cm@25℃
Quality confirms to the GB6682-2000 criteria of type Ⅲ water in laboratory application.
Ultrapure Water: Resistivity: 18.2 MΩ.cm@25℃, TOC<10ppb,  Microbe<1cfu/ml, Particle (>0.22 um)<1/ml
Application Range
HPLC, Atomic Absorption Spectroscopy, Mass Spectrometric Analysis, TOC Analysis, etc
Power
220V/50Hz;  50-100W
External Dimension
500×310×400 / 450×300×400 mm
450×300×400 mm
 
 
Model
P 05/ 10/ 20/ 40/ 60-B
Feed Water
City tap water,  feed pressure: 0.15-0.4Mpa, temperature: 5-40℃
Flow rate of
water production
5 / 10 / 20 / 40 / 60 L/hr
Flow rate of dispense water
1.5-2L/min
Output water parameter
Pure water: Resistivity ≥0.2MΩ.cm@25℃
Quality confirms to the GB6682-2000 criteria of type Ⅲ water in laboratory application.
Ultrapure Water: Resistivity: 2-15 MΩ.cm@25℃, TOC<20ppb, Microbe<1cfu/ml, Particle (>0.22 um)<1/ml
Application RangeTissue Detection, microbe detection, chemical analysis, blood testing, toxicity testing, environmental analysis, atomic absorption spectroscopy, students experiments,
Power
220V/50Hz;  50-100W
External Dimension
500×310×400 / 450×300×400 mm

Send your message to us: