FP6410 Flame Photometer
Application Areas
A lot of new technology and new materials are adopted. Such features as LCD display, menu-type instruction operation, curve storage, flameout protector and outside printout are enabled by the application of top grade embedded mono-chip computer circuit.
Specifications
◆RangeK:0~999.9;Na:0~999.9;Li:0~999.9
◆Sensitivity:K:0.2μg/mL;Na:0.2μg/mL
◆Linearity:K:0.2μg/mL;Na:0.7μg/mL
◆Drift:less than 3% within 15s
◆Reproducibility:3%
◆Aspiration rate:<6mL/min
◆Fuel supply:L.P.G.
Working conditions
◆Environmental Temperature: 10 oC~35oC
◆Relative humidity:≤85%
◆The instrument should be placed horizontally, avoiding direct sunshine radiation. There is no disturbance of strong electromagnetic field or vibration.
◆Fire extinguishing device should be equipped in the application field and good aeration should be guaranteed. Power supply: AC220V±22V, frequency 50Hz±1Hz,with good grounding.
Standard configuration:
◆Main Instrument 1
◆Power Cable 1
◆Chimney 1
◆Clearing rod 2
◆Air compressor 1
◆Printer SP-T 2425SK 1
Optional accessories:
◆Printer SP-T 2425SK
◆Air compressor
◆Lithium filter
Overall dimension & weight:
◆Shipping package dimension: 810-610-430mm
◆Gross weight: 23Kg
◆Net weight: 18Kg