MP522 Precision pH/Conductivity Meter
MP522 Precision pH/Conductivity Meter

MP522 Precision pH/Conductivity Meter

Brand:NANBEI

Model:MP522

Application:

product details

MP522 Precision pH/Conductivity Meter

水质多参数-竖版 (1) (1).jpg

Introduction

● There are three different measurement modes: stable display mode; time measurement mode; continuous measurement mode.
● Automatically recognize 13 kinds of pH buffer solution. User can choose anyone from four series of standard buffer solutions: Europe & USA series, NIST series, CHINA series and Self-defined series.
● High purified water and ammonia added purified water pH measuring mode can be set up, especially suitable for electric power and petrochemical industry.
● With EH redox potential measuring mode can directly display ORP value which against standard potential of hydrogen electrode.
● Automatically recognize 8 kinds of conductivity standard solution. User can choose anyone from two series of solutions: Europe & USA series and China series.
●Adopt advanced and original conductivity measurement technology, use 1 pc of conductivity electrode (K=1), it can meet measuring accuracy requirement for range within 0.5μS/cm ~ 200mS/cm by using only one-point calibration.
●Automatically select conductivity measurement range, when measuring high pure water, equipment automatically enter into temperature compensation of nonlinear model, especially for electricity, microelectronics and the pharmaceutical industry requirements.
● Meter has the only product serial number.
● Dustproof and waterproof meter meets IP54 standards. All sockets are seal protected by the silica gel caps.
●Meter meets with the requirement of international GLP standards. Its features up to automatic calibration, automatic temperature compensation, data storage, timing measurement, clock display, RS-232 output, function setting and self-diagnose etc.

Specifications

pH

Measuring range

Resolution

Accuracy

Input current

Input impendance

Stability

Temp. compensation range

(-1.999 ~ 19.999)pH

0.1/0.01/0.001 pH

Meter:±0.002 pH, connect meter with electrode:±0.01pH

≤1×10-12 A

≥3×1012 Ω

±0.002 pH/3h

(0 to 100) ℃(automatic or manual)

mV

Measuring range(mV/ORP/EH)

Resolution

Accuracy

-1999.9mV ~ 0 ~ 1999.9mV

0.1mV

±0.03% FS

Conductivity

Measuring range

Resolution

Accuracy

Temp. compensation range

Electrode constant

Reference temperature

Conductivity:(0.00~20.00)μS/cm (20.0~200.0)μS/cm

(200~2000)μS/cm  (2.00~20.00)mS/cm

(20.0~200.0)mS/cm (200~2000)mS/cm(K=10)

Resistivity:(0 ~ 100)MΩ·cm

TDS:  (0 ~ 100)g/L

Salinity: (0 ~ 100)ppt

0.01/0.1/1μS/cm  0.01/0.1/1 mS/cm

Meter:±0.5% FS,connect meter with electrode:±0.80% FS

(0 ~ 50)℃(automatic )

0.1 / 0.5 / 1 / 5 / 10 / 50 / 100  cm-1

25℃、20℃ and 18℃

Temperature

Measuring range

Resolution

Accuracy

-10℃~ 110℃

0.1℃

5~ 60℃range:±0.4℃  other range:±0.8℃

Other Technical Parameters:

Data storage

Storage content

Power

Communication connector

Size and weight

Quality and safety certification

300 groups

Measuring value series number, measuring value, temperature, ATC or MTC state, measuring date and time

DC9V/300mA

RS232

160 × 190 × 70mm/880g

ISO9001:2008, CE and CMC

Working Condition

Environment temperature

Environmental humidity

IP rating

5 ~ 35 ℃(0.01grade) 15 ~ 30 ℃(0.001grade)

≤75%

IP54 Dustproof and waterproof

MP522.jpg

Standard Kit Includes

1. Model MP522 Precision pH/Conductivity meter 1set
2. Model 602 flexible electrode holder 1pc
3. 201T-M plastic pH/ATC three-in-one combination electrode 1pc
4.2503-C glass pH combination electrode 1pc
5. MP500 temperature electrode 1pc
6. 2401-M conductivity electrode 1pc
7. pH standard buffer solution(pH4.00, pH7.00 and pH10.01/50mL) 1btl/each
8. 1413μS/cm conductivity standard solution(50mL) 2btls
9. 9V power adapter(meter use)e 1pc
10. RS-232 communication cable 1pc
11. MP522 communication software disk 1pc
12. Operation manual 1unit
13. Brief operation instruction 1unit
14. Packing List 1pc

water analysis instruments.jpg 